VH-1500 Respiratory Humidifer
Part Number
- 800-VH1500
Features
The product is used for humidifying breathing gas. There are 9 levels for temperature control for heating plate (from 45~75 ℃). That is to manually control the heating to humidification chamber for adjusting breathing gas temperature and humidification in order to let patient gain warm and humid breathing gas.
Specifications
Basic Specification
Dimension: 14.5 cm X 17 cm X 13.5 cm
Weight: 1.1 kg
Performance Specification
Mode: Non-heated Wire Mode (9-level heating control/temperature range: 45~75℃)
Flow: 5~40 LPM
Humidity Output: ≧33 mg/L
Warm-up Time: ≦30 min
Electrical Specification
Input Voltage: 100~120Vac /200~240Vac
Frequency: 50/60 Hz
Power Supply: 150 W
Thermal Cut-out: Manual reset cut out. Cut out temperature is 95+/-6℃ on heating plate.
Fuse: F2A/250Vac
Operation Environment
Temperature: 18~26 ℃
Pressure: 700~1060 hpa
Package: 6PCS/ BOX
Certification
◎ Passed IEC/EN 60601-1 A2: 2012 General Requirements for Basic Safety and Essential Performance
* General Requirements for Testing me Equipment
* Protection against Electronical Hazards from Me Equipment
* Protection against Mechanical Hazards of Me Equipment and Me Systems
* Protection against Unwanted and Excessive Radiation Hazards
◎ Passed IEC 60601-1-2:2015 Medical electrical equipment - Requirements and Tests
* Conducted Disturbance
* Radiated Disturbance
* Electrostatic Discharge (ESD)
* Electromagnetic Susceptibility (EMS)
* Electrical Fast Transient/burst (EFT)
* Surge
* Conducted Susceptibility (CS)
* Power Frequency Magnetic Field (PFMF)
* Voltage Dip, Interruption (DIP)
◎ Passed IEC 62304: 2015: Medical device software
* Muting Time
* No-sensor Detection
* No Heating Wire Detection
* Key Lock
* Status Record
* Control Indicator Display
* Control Key Operation
* Shut-down Alarm
◎ Passed ISO 80601-2-74: 2017 Particular Requirements for Basic Safety and Essential Performance
* Humidity Output Test
* Measured Gas Temperature
* Delivered Enthalpy
◎ Passed IEC 60529 Edition 2.1: 2001-IPX1